The integration of the Internet of Medical Things (IoMT) into healthcare systems has transformed patient care by enabling real-time monitoring, enhanced diagnostics, and enhanced operational efficiency. However, this increased connectivity has also expanded the attack surface for cybercriminals, raising significant cybersecurity and privacy concerns. This study focuses on the cybersecurity vulnerabilities of IoMT infusion pumps, which are critical devices in modern healthcare. Through a targeted literature review of the past five years, we analyzed seven current studies from a pool of 132 papers to identify security vulnerabilities. Our findings indicate that infusion pumps face vulnerabilities—such as device-level flaws, authentication and access control issues, network and communication weaknesses, data security and privacy risks, and operational or organizational challenges—that can expose them to lateral attacks within healthcare networks. Our analysis synthesizes findings from seven recent studies to clarify how and why infusion pumps remain vulnerable in each of these areas. By categorizing the security gaps, we highlight critical risk patterns and their implications. This work underscores the scope of the issue and provides a structured understanding that is valuable for healthcare IT professionals and device manufacturers. Ultimately, the findings can inform the development of targeted, proactive security strategies to better safeguard infusion pumps and protect patient well-being.

INTRODUCTION

The Internet of Medical Things (IoMT), a specialized subset of the Internet of Things (IoT), is revolutionizing the healthcare sector by enabling the seamless connection, interaction, and data exchange between medical devices and healthcare networks. According to the National Institute of Standards and Technology (NIST), IoT encompasses “the network of devices that contain the hardware, software, firmware, and actuators which allow the devices to connect, interact, and freely exchange data and information.” Within healthcare, IoMT devices include medical imaging systems, infusion pumps, wearable biosensors, and other connected tools that facilitate remote monitoring and enhance clinical workflows (Das, 2022a).

IoMT is transforming healthcare delivery by reducing costs, improving efficiency, and enabling remote patient monitoring. These advancements are

© 2025. Published by AHFE Open Access. All rights reserved. 47

48 Yener

particularly valuable for chronic disease management and rural healthcare access. Devices such as infusion pumps have become integral to daily clinical practices due to their ability to deliver fluids and medications with precision (O’Brien et al., 2018). However, the increasing reliance on IoMT also introduces significant cybersecurity risks. Infusion pumps, for instance, are vulnerable to a range of threats such as unauthorized access, data breaches, and malicious cyber-attacks that could jeopardize patient safety and interfere with healthcare operations.

Securing IoMT infusion pumps is essential for ensuring patient safety and preserving the reliability of healthcare systems. A compromised infusion pump can lead to life-threatening consequences by altering medication dosages or disrupting critical care workflows. Additionally, breaches involving Protected Health Information (PHI) expose healthcare providers to legal liabilities and reputational damage. Addressing these risks requires a comprehensive understanding of the vulnerabilities inherent in these devices and effective mitigation strategies tailored to their unique operational contexts.

RESEARCH OBJECTIVES This study aims to address an overview of existing research studies reported in recent publications by providing an analysis of cybersecurity vulnerabilities in IoMT infusion pumps. This paper particularly:

  • Identified five key vulnerabilities: Device-Level vulnerability, Authentication and Access Control Issues, Network and Communication vulnerability, Data Security and Privacy Risks, and Operational and Organizational Challenges.
  • Highlights how these vulnerabilities interconnect to expand attack surfaces within healthcare networks.

BACKGROUND

This work aims to provide context for the rapidly growing field of the Internet of Medical Things (IoMT) in healthcare, with a particular focus on infusion pumps. We summarize prior research regarding IoMT applications, highlight common security gaps in these devices, and discuss existing studies on IoMT security, emphasizing the research gap addressed in this paper.

IoMT Overview The IoMT market is rapidly expanding with estimates forecast that it will reach approximately $170 billion by 2030 (Grand View Research, 2023). Moreover, forecasts suggest that by 2026, smart hospitals around the globe will implement 7.4 million connected IoMT devices (averaging more than 3,850 devices per hospital) (Press 2022), with additional personal IoMT devices, such as smart wearables, contributing to further growth

Applications of IoMT span a range of healthcare functions. For example, these devices facilitate continuous monitoring of patients’ vital signs, enabling real-time data collection and prompt identification of potential health risks.

Threats and Security Strategies for IoMT Infusion Pumps 49

In elderly care, technologies such as smart belts utilize airbags to mitigate fall-related injuries (Tarbert & Singhatat, 2023). Similarly, Empatica’s smartwatch is used to detect early signs of seizures in individuals with epilepsy, alerting both patients and caregivers to enable timely intervention. Studies have demonstrated that such devices can achieve clinically acceptable levels of accuracy (Gerboni et al., 2023). Moreover, seamless integration of IoMT data into electronic health records (EHRs) boosts data accuracy and accessibility, leading to improved patient care and more informed clinical decision-making (Lutkevich n.d.).

Smart infusion pumps represent a critical component of IoMT in healthcare. These sophisticated devices incorporate features such as Dose Error Reduction Systems (DERS) and customizable drug libraries to improve the precision of medication delivery. Although infusion pumps are widely adopted in clinical settings, there are still challenges in optimizing their use through workflow integration, process improvements, and continued training (Hoffman & Bacon, 2020). Additionally, detailed descriptions of related devices for instance, insulin pumps reveal a blend of mechanical and electronic components that harmonize to ensure precise drug dosing (Singh et al., 2021).

Previous studies highlight risks from unsecured medical devices, such as unauthorized data access and interference with critical interventions (U.S. Department of Health & Human Services - Office for Civil Rights n.d., Proofpoint, 2024). Broader investigations show that the growing volume of data and interconnected systems in IoMT Increase threats of network overload and data breaches (Osama et al., 2023).

Ahmed et al. (2023) emphasize the need to improve data collection, protection, and storage to detect and mitigate threats. Despite technological advances, vulnerabilities persist including insecure electrocardiographs, compromised imaging systems, and unprotected dispensers (Vedere Labs, 2024). While these works address IoMT security generally, comprehensive categorization and targeted mitigation strategies for infusion pumps remain limited. Das et al. (2023) point the urgency of building robust security frameworks tailored to IoMT.

Known Vulnerabilities Despite the numerous benefits, IoMT devices, including smart infusion pumps face substantial cybersecurity challenges. Common vulnerabilities include weak encryption, poor authentication protocols, and inadequate network segmentation (Dzamesi & Elsayed, 2025). Many devices operate with outdated software or firmware and default credentials, leaving them exposed to unauthorized access and potential sabotage. One study examining connected devices identified over 4,000 vulnerabilities; approximately 2% of these were found in IoMT devices, with 80% considered critical (Labs, 2023).

Specifically for infusion pumps, known security gaps include unpatched software, insecure communication protocols, and potential physical-access exploits (Davis 2021; Nair 2022). (Das 2022b). Additionally, proper security measures such as robust authentication,

50 Yener

encryption, and regular firmware updates are often lacking, making these devices prime targets for cyber-attacks (Guarding the flow: Enhancing cybersecurity measures for infusion pump systems 2024).

Though infusion pumps offer clear clinical benefits, their cybersecurity challenges demand continued research. This paper contributes a focused analysis of infusion pump vulnerabilities and outlines security measures to secure IoMT infrastructure.

METHODOLOGY

Literature Search and Selection A comprehensive literature review was conducted to identify cybersecurity vulnerabilities specific to infusion pumps. We searched five academic databases: Scopus, Web of Science, Google Scholar, IEEE Xplore, and ACM Digital Library.

Search Strategy The search used the keywords “infusion pumps” and “vulnerability.” For IEEE Xplore and ACM Digital Library, we applied “infusion pumps” and “security vulnerability,” limiting results to 2020–2025. In Google Scholar: “infusion pumps” AND “security vulnerability” AND “IoMT” (“smart infusion pump” OR “drug infusion system”) AND (“cyber threat” OR “hacking” OR “attack surface”) AND (“healthcare” OR “medical device”), also within the 2020–2025 range. ACM Digital Library was searched using: [Abstract: “infusion pump”] OR [Abstract: “insulin pump”] AND [Abstract: vulnerability]. IEEE Xplore was queried with: (“All Metadata”: infusion pump) AND (“All Metadata”: vulnerab*) AND (“All Metadata”: cybersecurity). Scopus and Web of Science were both searched using: (“infusion pumps” AND “vulnerability”).

Inclusion and Exclusion Criteria In Scopus, terms were applied to the “Article Title, Abstract, and Keywords” fields; in Web of Science, within the “Topic” field. All databases were filtered for publications from 2020–2025.

Included studies were peer-reviewed articles and conference papers explicitly addressing cybersecurity vulnerabilities in infusion pumps. Excluded materials included: Dissertations, Irrelevant papers (e.g., clinical studies without a security focus), Documents without full-text access, Duplicate entries, Grant announcements or abstracts lacking technical detail, Review papers.

Threats and Security Strategies for IoMT Infusion Pumps 51

Final Selection From an initial pool of 132 papers, seven met the selection criteria after rigorous screening. Table 1 summarizes the search results from each database.

Table 1: Search results from various databases.

Table available in the original PDF paper.

Relevance Selected

Database Search Results

Papers Scopus 28 5 1 Web of Science 18 4 1 Google Scholar 77 17 3 IEEE Xplore 6 3 1 ACM Digital Library 3 1 1 Total 132 30 7

Data Extraction and Analysis The analysis focused on identifying the most commonly reported vulnerabilities. The following five security issues emerged as the primary concerns discussed in the literature: device-level flaws, authentication weaknesses, network vulnerabilities, data privacy risks, and operational challenges.

FINDINGS This section describes the main outcomes of our review, which found five key cybersecurity vulnerabilities in IoMT infusion pumps, summarized below and in Table 2.

Device-Level Vulnerabilities A prominent issue is buffer overflow, where malicious packet injection bypasses memory protections, enabling arbitrary code execution. Beyrouti et al. (2024) demonstrated this via simulated attacks on medical IoT gateways, causing system crashes and data corruption. Similarly, sensor manipulation attacks, as shown by Chang et al. (2023), exploit adversarial inputs to alter glucose readings in artificial pancreas systems, leading to incorrect insulin dosing. Stergiopoulos et al. (2023) further validated these risks by illustrating how manipulated sensor data disrupts PID controller outputs in insulin pumps. Physical tampering and legacy systems exacerbate these risks. Taylor et al. (2022) documented physical exploits, such as replacing WiFi cards with malicious CompactFlash components (CVE-2016- 8375), which bypass digital safeguards entirely. Such findings underscore the urgent need for hardware integrity checks and firmware updates to mitigate life-threatening risks.

Authentication and Access Control Issues Authentication and access control vulnerabilities in infusion pumps, as reported in the reviewed studies, expose critical risks through weak credential

52 Yener

management and insufficient privilege enforcement. These flaws enable unauthorized access to device settings or sensitive patient data, often stemming from design oversights in device firmware and communication protocols.

A recurring issue is the use of hard-coded credentials, which manufacturers embed into device firmware for administrative access. Taylor et al. (2022) identified static passcodes in Alaris infusion pumps that granted unrestricted access to network configurations, a flaw exploited in real-world scenarios to manipulate device parameters. Similarly, Beyrouti et al. (2024) documented cases where missing authentication checks allowed attackers to bypass critical functions, such as firmware updates, through unvalidated input channels.

Denial-of-Service (DoS) attacks further compound these risks by disrupting sensor functionality. Chang et al. (2023) demonstrated adversarial attacks on Continuous Glucose Monitoring (CGM) sensors, where maxed- out readings caused temporary system failures. Griffy-Brown et al. (2022) corroborated this by highlighting code injection vectors that exploit weak authentication to overload device resources. Such attacks not only impair device availability but also create entry points for lateral movement within healthcare networks.

Network and Communication Vulnerabilities Network and communication vulnerabilities in infusion pumps arise from insecure protocols and misconfigurations that expose devices to data interception or remote exploitation. These weaknesses, reported in the reviewed literature, highlight systemic risks in healthcare networks due to the interconnected nature of IoMT devices. For example, a critical issue is clear text data transmission, where sensitive information, such as sensitive data configurations or patient data, is exchanged without encryption. Beyrouti et al. (2024) demonstrated this vulnerability through simulated attacks on medical IoT gateways, showing how unencrypted communication channels allow adversaries to intercept and manipulate data via man-in-the-middle (MITM) attacks. Likewise, Mejía-Granda et al. (2024) identified FTP server exploitation in syringe pumps, where unsecured file transfer protocols enabled unauthorized access to device settings or firmware. The consequences of such vulnerabilities extend beyond individual devices. For instance, compromised infusion pumps can serve as entry points for lateral movement within hospital networks, enabling attackers to breach Electronic Health Records (EHRs) or disrupt clinical workflows. Taylor et al. (2022) further emphasized risks like credential theft via physical network card replacement (CVE-2016-8375), illustrating how hardware-level flaws compound network insecurities.

Data Security and Privacy Risks Data security and privacy risks in infusion pumps involve exposure or unauthorized access to protected health information (PHI), as reported across multiple studies. Beyrouti et al. (2024) demonstrated how

Threats and Security Strategies for IoMT Infusion Pumps 53

unencrypted communication between infusion pumps and IoT gateways allows adversaries to intercept sensitive data via man-in-the-middle (MITM) attacks. Similarly, Mejía-Granda et al. (2024) identified sensitive information leakage in syringe pumps, where unsecured FTP protocols exposed PHI, including patient identifiers and treatment details. Taylor et al. (2022) documented hard coded passwords in Alaris pumps that granted unrestricted access to network settings, allowing attackers to exfiltrate PHI or manipulate device parameters and show that insufficiently protected credentials further exacerbate these risks. These vulnerabilities, stemming from insecure data handling practices, threaten patient confidentiality and institutional compliance with healthcare regulations.

Operational and Organizational Challenges Operational and organizational challenges in securing infusion pumps stem from systemic issues such as outdated software, misconfigurations, and inadequate maintenance practices, as reported in the reviewed studies. Mejía- Granda et al. (2024) identified outdated operating systems (e.g., Windows XP) in medical devices, which lack modern security features and leave pumps exposed to unpatched vulnerabilities. The study also highlighted security misconfigurations, such as improper network segmentation and disabled encryption policies, which enable attackers to bypass safeguards. For instance, Taylor et al. (2022) documented maintenance mode vulnerabilities in Alaris pumps, where unprotected diagnostic interfaces allowed unauthorized access to error logs and network configurations. These challenges create persistent vulnerabilities that adversaries can exploit over extended periods.

DISCUSSION

Despite growing awareness, many of the same fundamental weaknesses persist across studies and device models. A consistent pattern emerges: vulnerabilities span multiple layers from insecure firmware and default passwords to exposed networks and weak governance. Poor authentication practices stand out in nearly every source. Both targeted case studies by Taylor et al. (2022) and broad surveys by Mejía-Granda et al. (2024) report hard-coded or default credentials, signalling a systemic failure to address even basic access controls. Similarly, the frequent use of outdated software and unpatched flaws points to an industry lag in keeping medical devices secure against evolving threats. When one study finds 75% of infusion pumps have known vulnerabilities and another confirms active CVEs on widely used models, it’s clear that these are not isolated issues but part of a broader, ongoing challenge.

Another recurring problem is insecure communication and integration. Several studies identify unencrypted channels, open ports, and flat network configurations as major risks as shown by Taylor et al. (2022) & Mejía- Granda et al. (2024). These network gaps often connect directly to other vulnerabilities: weak credentials can be exploited remotely due to open access, leading to data exposure and long-term compromise. The reviewed

54 Yener

literature, spanning technical tests and policy analysis—shows how these issues reinforce one another. Even advanced devices with AI features, such as those discussed by Chang et al. (2023) and Stergiopoulos et al. (2023), remain vulnerable if core weaknesses like access control are not addressed Chang et al. (2023) and Stergiopoulos et al. (2023).

Overall, infusion pump cybersecurity demands a coordinated approach. The studies reviewed recommend building security into device design, enforcing strong authentication and encrypted communication, segmenting networks, and promoting proactive risk management in clinical settings (Taylor et al., 2022; Mejía-Granda et al., 2024; Beyrouti et al. 2024). By categorizing vulnerabilities across technical and organizational dimensions, this review connects isolated findings into a more complete picture. A weakness identified in one context often echoes broader trends across the field. By drawing out these connections, this synthesis helps inform more holistic strategies for improving the security of these critical healthcare devices.

Limitations One of the limitations of our study is relying on a small number of recent papers. Although these papers are insightful it may not encompass the full range of perspectives in infusion pump vulnerabilities. Key perspectives from industry technical reports or a wider variety of device types is not included. As a result, our brief review and synthesized findings does skew towards the particular vulnerabilities and solutions emphasized in these selected works. Additionally, We focused primarily on infusion pumps and closely related hospital devices. This narrower focus, while keeping the analysis on track, limits the breadth of our implications.

CONCLUSION

Our review shows that IoMT infusion pumps face a broad set of cybersecurity vulnerabilities across five main areas: device-level flaws, authentication and access control weaknesses, network and communication issues, data security risks, and operational shortcomings. These vulnerabilities often intersect poor access control can open paths for network intrusions that compromise device behavior or leak sensitive information. The consequences are serious: a compromised pump could alter medication delivery or serve as a foothold for broader attacks within healthcare systems. To reduce these risks, organizations need to implement layered defenses, such as removing hard-coded credentials, maintaining up-to-date software, securing communication channels, and isolating devices from general networks. These technical efforts must be backed by institutional practices like staff training and integration of device security into broader risk management. The findings across the seven studies underscore the need for both technical and organizational action. As infusion pumps remain vital to patient care, securing them should be considered a core safety concern. Continued collaboration among manufacturers, healthcare stakeholders, and regulators will be key to strengthening standards, improving designs,

Threats and Security Strategies for IoMT Infusion Pumps 55

and ensuring these devices are resilient in today’s connected healthcare environments.

REFERENCES

Ahmed, S. et al. (2023), ‘Insights Into Internet of Medical Things (IoMT): Data fusion, security issues and potential solutions’, Information Fusion 102, 102060. URL: https://doi.org/10.1016/j.inffus.2023.102060. Beyrouti, M., Lounis, A., Lussier, B., Bouabdallah, A. & Samhat, A. E. (2024),

‘Vulnerability-oriented risk identification framework for IoT risk assessment’, Internet of Things 27, 101333. Chang, P., Krish, V. & Rahmati, A. (2023), Security analysis of rl-based artificial pancreas systems, in ‘Proceedings of the 2024 Workshop on Cybersecurity in Healthcare’, pp. 69–76. Das, A. (2022a), ‘Infusion pump vulnerabilities: Common security gaps’. URL:

https://unit42.paloaltonetworks.com/infusion-pump-vulnerabilities/. Das, A. (2022b), ‘Know your infusion pump vulnerabilities and secure your healthcare organization’. Accessed: 22 February 2025. URL: https:// unit42.paloaltonetworks.com/infusion-pump-vulnerabilities/. Das, S. et al. (2023), Emerging cyber threats in healthcare: A study of attacks in IoMT ecosystems, in ‘2023 7th International Conference on I-SMAC (IoT in Social, Mobile, Analytics and Cloud) (I-SMAC)’, pp. 77–82. URL: https://doi.org/ 10.1109/I-SMAC58438.2023.10290147. Davis, J. (2021), ‘Report: Direct patient safety risk posed by infusion pump

vulnerability exploit’. Accessed: 21 March 2025. URL: https://www.scworld.com/ analysis/report-direct-patient-safety-risk- posed-by-infusion-pump-vulnerability- exploit. Dzamesi, L. & Elsayed, N. (2025), ‘A review on the security vulnerabilities of the IoMT against malware attacks and ddos’, arXiv preprint arXiv:2501.07703. URL: https://arxiv.org/abs/2501.07703. Gerboni, G. et al. (2023), ‘Prospective clinical validation of the empatica embraceplus

wristband as a reflective pulse oximeter’, Frontiers in Digital Health 5, 1258915. Grand View Research (2023), ‘Internet of things in healthcare market size report, 2030’. Accessed: 23 February 2025. URL: https://www.grandviewresearch.com/ industry-analysis/internet-of-things-iot-healthcare-market. Griffy-Brown, C., Miller, H., Chun, M. & Johnson, K. (2022), Cyber risk case

analysis in wearables and medical devices: Developing a cyberbio security risk framework, in ‘2022 Portland International Conference on Management of Engineering and Technology (PICMET)’, IEEE, pp. 1–6. Guarding the flow: Enhancing cybersecurity measures for infusion pump systems (2024), Healthcare IT News. Accessed: 26 February 2025. URL: https://www.healthcareitnews.com/news/guarding-flow-enhancing- cybersecurity-measures-infusion-pump-systems. Hoffman, L. & Bacon, O. (2020), Infusion Pumps, Agency for Healthcare Research and Quality (US). Accessed: 26 February 2025. URL: https:// www.ncbi.nlm.nih.gov/books/NBK555506/. Islam, S., Abba, A., Ismail, U., Mouratidis, H. & Papastergiou, S. (2022),

‘Vulnerability prediction for secure healthcare supply chain service delivery’, Integrated Computer-Aided Engineering 29(4), 389–409. Labs, V. (2023), ‘The 5 riskiest connected devices in 2023: It, IoT, Ot, IoMT’.

Accessed: 24 February 2025. URL: https://www.forescout.com/blog/riskiest- connected-devices-it-iot-ot-iomt/.

56 Yener

Lutkevich, B. (n.d.), ‘What is the internet of medical things (IoMT)? — definition from techtarget, search IoT’. Accessed: 24 February 2025. URL: https:// www.techtarget.com/iotagenda/definition/IoMT-Internet-of-Medical-Things. Mejía-Granda, C. M., Fern’andez-Alem’an, J. L., Carrillo-de Gea, J. M. & Garc’ıa-

Bern’a, J. A. (2024), ‘Security vulnerabilities in healthcare: an analysis of medical devices and software’, Medical & biological engineering & computing 62(1), 257–273. Nair, P. (2022), ‘Security gaps in smart infusion pumps risk patient data’. Accessed: 21 March 2025. URL: https://www.govinfosecurity.com/security-gaps-in-smart- infusion-pumps-risk-patient-data-a-18653. O’Brien, G., Edwards, S., Littlefield, K., McNab, N., Wang, S. & Zheng, K. (2018),

Securing wireless infusion pumps in healthcare delivery organizations, Technical Report 1800-8, National Institute of Standards and Technology (NIST). URL: https://doi.org/10.6028/NIST. SP.1800–8. Osama, M. et al. (2023), ‘Internet of medical things and healthcare 4.0: Trends, requirements, challenges, and research directions’, Sensors (Basel, Switzerland) 23(17), 7435. URL: https://doi.org/10.3390/s23177435. Proofpoint (2024), ‘2024 ponemon healthcare cybersecurity report’. Singh, N. et al. (2021), ‘A formal approach to rigorous development of critical

systems’. Smart Hospitals to Deploy Over 7 Million Internet of Medical Things — Press (2022), Juniper Research. Accessed: 23 February 2025. URL: https:// www.juniperresearch.com/press/smart-hospitals-to-deploy-over-7mn-iomt/ Stergiopoulos, G., Kotzanikolaou, P., Konstantinou, C. & Tsoukalis, A. (2023),

‘Process-aware attacks on medication control of type-i diabetics using infusion pumps’, IEEE Systems Journal 17(2), 1831–1842. Tarbert, R. & Singhatat, W. (2023), ‘Real world evidence of wearable smartbelt for mitigation of fall impact in older adult care’, IEEE Journal of Translational Engineering in Health and Medicine 11, 247–251. Taylor, K., Smith, A., Zimmel, A., Alcantara, K. & Wang, Y. (2022), Medical device

security regulations and assessment case studies, in ‘2022 IEEE 19th International Conference on Mobile Ad Hoc and Smart Systems (MASS)’, IEEE, pp. 742–747. U.S. Department of Health & Human Services - Office for Civil Rights (n.d.), Online resource. Accessed: 2 December 2024. Vedere Labs (2024), ‘The riskiest connected devices 2024’.

Cover of the proceedings volume Human Factors in Cybersecurity

Explore the complete digital volume

Human Factors in Cybersecurity

Read this individual research paper free through the original publisher, or explore the full digital proceedings book.

ISBN 978-1-964867-44-1

Explore this ISBN in the AHFE Bookstore Listed in the July 2026 bookstore catalogue; search by ISBN to confirm current availability · Digital book purchase optional

Explore Related Research

Related research 8 matches

More papers from this volume

← Back to Research ExplorerBrowse the full library →